我县医疗单位30mAX光机均系70年代产品。防护条件差,1.部份X光机无铝过滤或厚度不够,导致球管出线口输出量增加。2.屏周挂帘和屏侧铅橡皮长度和宽度不够。3.聚光筒铅当量不够,对上述问题,采取了如下措施。 1.限光、提质、准直。用2mm铅板自制古钱板30个,每台机配备一个,该板小孔制作根据使用时焦皮距不同,保证投照到荧光屏上时周边至少有1cm无光区。用2mm铝板制作过滤片30个,起到滤除低能射线,以提高线束质量。还在收光器前加聚光筒,并以2mm铁皮衬内。2.扩大荧光屏,加
My county medical unit 30mAX light machine are the 70’s products. Poor protection conditions, 1. Part of the X-ray machine without aluminum filter or thickness is not enough, leading to increased output tube outlet. 2.Screen weeks curtain and side of the lead rubber length and width is not enough. 3 condenser tube lead equivalent is not enough, the above problems, has taken the following measures. 1. Limit light, improve quality, collimation. Made with 2mm clickers 30 antique coins, each with a machine, the board made of small holes according to the use of different focal lengths, to ensure that when projected to the screen when the surrounding area of at least 1cm matte. Made with 2mm aluminum filter 30, played a filter low-energy rays, in order to improve the quality of the harness. Still in front of the condenser plus condenser, and 2mm metal lining. 2. Expand the screen, plus