,The effect of pitch and collimation on radiation dose in spiral CT

来源 :核技术(英文版) | 被引量 : 0次 | 上传用户:a_hai1983
下载到本地 , 更方便阅读
声明 : 本文档内容版权归属内容提供方 , 如果您对本文有版权争议 , 可与客服联系进行内容授权或下架
论文部分内容阅读
Measurements of radiation dose to patients in spiral computed tomography (CT) were completed for various collimations, table speeds and pitch. A standard CT head dosimetry phantom and thermoluminescent dosimeters (TLD) were used for the measurement. The.effect of collimation and pitch on radiation dose was studied. The results indicated that the radiation dose at the given tube current, voltage and rotation speed was inversely proportional to pitch. And the increasing times of dose were as decreasing times of pitch. This regular pattern was tenable for radiation dose at both central holes and peripheral holes of the phantom at pitch = 1, >1 and <1. The collimation had no impact on the radiation dose. The results also indicated that radiation dose at central holes was nearly equal to that at peripheral holes. There was no significant difference between them statistically. The study demonstrates that the pitch in spiral CT scans is the primary parameter and has significant impact on radiation dose.