宣城市人民医院于2013年引进了飞利浦Allura Xper FD20端血管造影X线摄像系统用于介入治疗。Allura Xper FD系统在运行过程中,会出现Bodyguard警示信息,同时机架运动速度变慢,并伴随有蜂鸣器警示声,影响介入手术的进程。设备开机后,显示器显示如下信息:“Reduced speed,bodyguard blocks high speed movement”“Warning:bodyguard faulty,move at own risk Clean bodyguard”“Warning:bodyguard
Xuancheng People’s Hospital introduced the Philips Allura Xper FD20 angiography X-ray system in 2013 for interventional therapy. Bodyguard alerts appear during operation of the Allura Xper FD system, and the rack moves slowly and is accompanied by a buzzer warning that affects the interventional procedure. After the device is powered on, the display shows the following message: ”Reduced speed, bodyguard blocks high speed movement “ ”Warning: bodyguard faulty, move at own risk Clean bodyguard “ ”Warning: bodyguard