我们对22例正常人(440眼)先后进行Goldmann和Schiotz眼压计对比测量,统计学发现有显著性差异.Goldmann眼压计较Shiotz眼压计测量结果低〔平均值低0.281kPa(2.11mmHg)〕,结果也较Schiotz眼压计测量准确.
We have 22 cases of normal subjects (440 eyes) have Goldmann and Schiotz contrast tonometry, statistically significant differences were found .Goldmann Tonometer than Shiotz Tonometer measurement results [average low 0.281kPa (2.11mmHg) 〕, The result is also more accurate than the Schiotz tonometer.