论文部分内容阅读
例1 故障现象:通电后踩动脚踏开关,压力表指示压力偏低,调节压力调节旋钮变化不大。分析检修:牙钻手机正常工作所需压力应达到0.2~0.25 MPa,压力过低手机车针旋转速度偏低则钻不动牙。出现压力偏低的原因主要有:空压机供气量不足、管路漏气等。从空气压缩机转动时发出的声音判断好像是负载变轻。首先检查电动机与空压机气泵之间的传动皮带并未松弛,检查气、水管路及管线各接头处也无漏气现象。估计此故障由空压机自身故障引起。经检查发现空压机油标视孔处全是油污,看不清真正油位,气泵尾部的油水分离器内也全是油污。将油水分离器下方的排污螺丝堵拧掉也没有液体流出,并且发现此螺丝堵已松动,即使把空压机开动起
Example 1 Symptom: After the treadle is energized, the pressure gauge indicates the pressure is low, and the adjustment pressure adjustment knob does not change much. Analysis and maintenance: dental drill phone pressure required for normal work should reach 0.2 ~ 0.25 MPa, the pressure is too low mobile phone needle rotation speed is too low to drill fixed teeth. The reasons for the low pressure are: lack of air compressor supply pipe leakage and so on. The sound from the air compressor turns out to be a lighter load. First check the motor and air compressor air pump belt between the transmission is not loose, check gas, water pipes and pipe joints at no leak. This fault is estimated to be caused by the air compressor itself. The inspection found that the air compressor oil standard hole is full of oil, can not see the real oil level, the end of the oil-water separator is also full of oil. The oil and water separator below the sewage screw plug screwed off no liquid out, and found that the screw plug has been loose, even if the air compressor to start