目的了解2013年哈尔滨市放射诊断设备性能状况,以便开展放射诊断设备的防护,保证医患人员的身心安全。方法按照国家技术规范与防护标准规定对37台X射线摄影设备和17台X射线透视设备进行性能检测。结果 37台X射线摄影设备中有29台检测项目全部合格,合格率为78.4%;17台X射线透视设备中有14台检测项目全部合格,合格率为82.4%。结论哈尔滨市放射诊断设备性能大部分都符合国家有关规定,有少数不合格设备需要进一步整改,为放射诊断的质量和安全提供可靠的保证。
Objective To understand the performance status of radiation diagnostic equipment in Harbin in 2013 in order to carry out radiation diagnostic equipment protection and ensure the medical and psychological safety of medical staff. Methods The performance of 37 X-ray radiography equipment and 17 X-ray radiographic equipment were tested according to the national technical regulations and protection standards. Results Of the 37 X-ray equipment, 29 were all qualified with a pass rate of 78.4%. Of the 17 X-ray equipment, 14 were all qualified with a pass rate of 82.4%. Conclusion Most of the performance of diagnostic radiology equipment in Harbin are in compliance with the relevant regulations of the state. There are a few unqualified equipment that need further rectification to provide a reliable guarantee for the quality and safety of radiological diagnosis.