Objective To compare the effect of rapid biological indicator and conventional biological indicator on the monitoring of ethylene oxide sterilization, and to find a rapid detection method of ethylene oxide sterilization. Methods The ethylene oxide biological indicator resistance evaluation system was used to compare the two kinds of biological indicators to monitor the sterilization effect of ethylene oxide. Results In the same conditions of ethylene oxide sterilization, sterilization 20 min and 120 min, the two methods of biological indicators of biological culture consistent results; 70 min sterilization period, the two culture methods for 48 h sustained the same results, However, the results of the rapid culture method at 4 h were not consistent with the results at 48 h. Conclusion Compared with the conventional culture method, ethylene oxide biological monitoring rapid culture method has the ability of more rapid evaluation of sterilization effect, but there may be false positive results.